rotasept®

rotasept® is a ready-to-use solution for the disinfection and cleaning of rotary instruments

Our Plus

  • comprehensively effective
  • highly suitable for use in ultrasonic bath
  • aldehyde-free
  • ready-to-use

Downloads

Fields of Application

For manual cleaning and disinfection of burs, cutters, diamond grinders and other rotating precision instruments. Due to its excellent material compatibility and comprehensive efficacy, rotasept® provides all important product properties for reprocessing rotating precision instruments as well as instruments made of steel and stainless steel.

Product Profile


Instructions for use

For professional use only. rotasept® is a ready-to-use product and is added undiluted to the bur bath.
If rotasept® is used as cleaner, place each small rotating instrument in the bur bath solution immediately after use in order to avoid surface-drying of dirt. Rinse all instruments thoroughly after reprocessing, using water of at least drinking water quality or preferably sterile distilled water or fully deionised water, to completely remove residues of the solution.

Immersion time for final disinfection: 15 minutes. Longer contact times should be avoided to ensure best material compatibility. Replace working solutions every working day and if contamination is clearly visible. National regulations may require that cleaning and final disinfection are carried out in two separate process steps. Do not mix with cleaning agents and disinfectants. Please follow the reprocessing recommendations of the instrument manufacturers.

Material compatibility:
Alkali- and alcohol-sensitive plastics, aluminium parts and instruments which have been made using synthetic resin adhesives should not be placed in rotasept®. We recommend gigasept® instru AF for these instruments.

Ultrasonic bath: rotasept® is highly suitable for the use in ultrasonic bath.

Microbiological efficacy

EfficacyConcentrationContact time
bactericidal
EN13727, EN14561, in accordance with VAH
- dirty conditions
ready-to-use5 min.
mycobactericidal
EN14348, EN14563, in accordance with VAH
- dirty conditions
ready-to-use15 min.
yeasticidal
EN13624, EN14562, in accordance with VAH
- dirty conditions
ready-to-use5 min.
fungicidal
EN13624, EN14562
- dirty conditions
ready-to-use5 min.
virucidal
EN14476
- dirty conditions
ready-to-use1 min.
virucidal
in accordance with DVV (German Registered Association for Combating Viral Diseases
ready-to-use5 min.

Certificates

  • VAH certificate
  • IHO listed

Product data

Composition:
100 g concentrate contains the following active substances: 1.6 g Potassium hydroxide, 10.0 g Propan-2-ol, 0.1 g 2-Ethylhexanol. labelling according to Regulation (EC)648/2004: < 5 % non-ionic surfactants.



Chemical-physical data

Flash point:36 °C / Method : DIN 51755 Part 1
Colour:colourless
Form:liquid
Density:ca. 1,00 g/cm3 / 20 °C
pH:13,7 / 20 °C

Special advice

Always read the label and product information before use.
Any serious incidents related to the product are to be reported to the manufacturer and competent authority. Keep container tightly closed. Protect from frost, heat and sunlight. Store in the original container. Do not use the product after the expiry date.

Information for order

ItemDelivery formItem no.
rotasept® 2 L5/Cartonon request

These products are not available in every country. For more information please contact our local subsidiary or distributor.

Environmental information

 schülke manufactures products economically and with advanced, safe and environmentally friendly production processes while at the same time maintaining out high quality standards.

EuroNormCE 0297

Privacy Policy

We use analysis methods (e. g. cookies) to measure how often our site is visited and how it is used.
We embed third-party content from other providers (e. g. videos) on our site. We have no influence on further data processing and any tracking by the third party provider.
In this context, we also use service providers in third countries outside the EU that do not have an adequate level of data protection, which harbors the following risks: Access by authorities without informing the data subject, no data subject rights, no legal remedies, loss of control.
With your settings you consent to the processes described above. You can revoke your consent with effect for the future. You can find more information in our privacy policy.