Fields of Application
Instructions for use
gigasept® FF (new) is a concentrate and is diluted with cold water to the desired working concentration.
Application example:
1 litre of a 6 % working solution corresponds to 940 ml of water and 60 ml of gigasept® FF (new).
Immerse endoscopes and instruments to be reprocessed in the gigasept® FF (new) solution. Ensure complete covering of the surfaces, also of hollow instruments, and allow them to soak in the solution. After disinfection, thoroughly rinse all instruments, using water of at least drinking water quality, or preferably sterile distilled water or fully deionised water, to completely remove any residues of the solution. Please follow the reprocessing recommendations of the instrument manufacturers.
Standing time:
A standing time of 14 days applies if the working solution is only used for the disinfection of previously cleaned instruments. National regulations may limit the timespan during which the solution can be used.
Apart from the manual treatment of endoscopes and instruments, gigasept® FF (new) is also suitable for semiautomated machines that operate by a cyclic process at room temperature.
Certificates / Norms



Product data
Composition:
100 g solution contain the following active substances: 93.9 g Reaction product of DMO-THF, ethanol and water.
Chemical-physical data
Special advice
Always read the label and product information before use. For professional use only. Any serious incidents related to the product are to be reported to the manufacturer and competent authority.
Material compatibility:
metals, rubber, glass, plastics, flexible endoscopes.
Interactions:
As a general rule, do not mix gigasept® FF (new) with other products. If gigasept® FF (new) comes into contact with application solutions containing amines, there may be precipitation and irreversible discolorations, for example if cleaning agents containing amines are used during precleaning and are not adequately rinsed off.
Storage:
Do not store for long periods above 25 °C. Keep container tightly closed. Protect from frost, heat and sunlight. Store in the original container. Do not use the product after expiry date.
Information for order
Item | Delivery form | Item | GTIN |
---|---|---|---|
gigasept FF neu -INT- 2 l FL | 5 / carton(s) | 70003414 | 4032651984367 |
gigasept FF neu -INT- 5 l KA | 1 / canister(s) | 70003415 | 4032651984381 |