Source of supply
Fields of Application
Before applying thermosept® NKP, make sure that the items to be cleaned and reprocessed are acid-resistant.
thermosept® NKP is suitable for use in all common washer/disinfectors and multichamber waher/disinfectors.
Instructions for use
Dosage:
For neutralisation after alkaline cleaning: 0.1 - 0.2 % (1 - 2 ml/l). Dosing is performed by machine-integrated dosing pumps.
Basic cleaning of instruments and machine interiors:
- as an acidic pre-cleaner: 0.2 - 2 % (2 - 20 ml/l)
- basic cleaning of instruments and machine interiors: approx. 4 %, (40 ml/l)
- as a cleaner in a manual immersion bath: approx. 4 %, (40 ml/l)
thermosept® NKP is not suitable for coloured metals, chromium plated parts and anodized aluminium.
Please make sure that the items to be cleaned, the machine and the drains are acid resistant.
Using demineralised water for final rinsing is recommended and improves the rinsing result.
thermosept® NKP can be used in washer/disinfectors and multi-chamber washer/disinfectors.
Please note recommendations of machine and instrument manufacurers.
Certificates / Norms

Product data
Composition:
Contains orthophosphoric acid.
Special advice
Always read the label and product information before use. For professional use only. Any serious incidents related to the product are to be reported to the manufacturer and competent authority. Please note recommendations of machine and instrument manufacturers. Keep container tightly closed. Protect from frost, heat and sunlight. Store at room temperature in the original container. (Storage temperature: 5 - 25 °C).
Information for order
Item | Delivery form | Item | GTIN |
---|---|---|---|
thermosept NKP 6 kg KA | 1 / canister(s) | 146403 | 4032651464036 |