Product Picture Master

gigasept® PAA

Ready-to-use high level disinfectant based on peracetic acid for the disinfection of flexible endoscopes and medical instruments. gigasept® PAA is a medical device class IIb.

Our Plus

  • Aldehyde-free
  • Ready to use
  • Effective against bacteria, fungi, viruses, spores
  • No activation necessary

Downloads

Fields of Application

Manual or semi-automated disinfection/processing of pre-cleaned flexible endoscopes, ophtalmological equipment, surgical instruments made of stainless steel and heat sensitive medical devices.

Instructions for use

Use disinfectants safely. Always read the label and product information before use. For professional use only. Read SDS before use.
gigasept® PAA is a ready to use product. gigasept® PAA is active immediately and stable over 7 days or 50 cycles of disinfection. The peracetic acid (PAA) concentration in solution can be tested using Schülke PAA test strips. In case of visible contamination, the bath should be renewed.
Always ensure the pre-cleaning of the endoscope and medical instruments and a complete wetting of the surfaces.
Immerse hollow objects in such a way that air escapes completely from the lumen.
After the necessary contact time rinse carefully with sterile or filtered water before re-use. Do not exceed the recommended contact time.
A located ventilation over the bath is recommended. Cover the bath of gigasept® PAA as often as possible.
Do not dilute gigasept® PAA. Do not mix with other products (cleaner, disinfectants, etc.), particularly with aldehydes or derivates of halogens (e.g. glutaral, sodium hypochlorite, etc.). Do not use after expiry date.


Certificates / Norms

0297

Product data

Composition: 100 g solution contains the following active substances: 0,12 g Peracetic acid [CAS n°79-21-0].

Chemical-physical data

Color
colourless
Density
ca. 1,02 g/cm3 / 20 °C
Flash point
> 100 °C
Form
liquid
pH
ca. 3,5
Viscosity, dynamic
not determined
Odor
vinegar-like

Special advice

Safety:
WARNING - irritant product- always refer to the product label and the Safety Data Sheet before use.
gigasept® PAA is not classified as a dangerous formulation by ingestion (Test OECD 401) or irritation to skin (OECD 404).
gigasept® PAA does not generate toxic vapour - the average exposure limit of the ingredients is respected.
gigasept® PAA is not sensitising (Test OECD 406). Gloves must be worn when handling disinfectants not intended for the skin.
Any serious incident that has occurred in relation to this device should be reported to Bioxal and your competent authority.

Compatibility:
gigasept® PAA can be used in semi-automated endoscope reprocessors that are compatible with peracetic acid. Please check the compatibility of endoscope and reprocessors with paracetic acid.
Incompatible materials: iron, toolsteel, aluminium, platinium, nickel -/ chromeplated or non ferrous metal alloys parts (e.g. cooper and brass).
Compatible materials: (limited to the maximum contact time; obligatory rinse after contact) polycarbonate, polyurethane, silicone, polysulfone, polyamide, polyester, polystyrene.
Storage and disposal:
Store and use only in a well ventilated place. Store at ambient temperature in the original canister. Always close canister tightly after removal of product. Canister with venting cap. Do not keep the venting cap sealed. Store in upright position only. Protect from heat and sunlight. Do not store with alkaline products. Never return used product into the original canister. Do not reuse empty container. Dispose of the product including packaging as hazardous waste in compliance with local and national regulations.

Information for order

ItemDelivery formItemGTIN
gigasept PAA 5 l KA1 / canister(s)1957514032651957514

Environmental information

schülke manufactures products economically and with advanced, safe and environmentally friendly production processes while at the same time maintaining our high quality standards.

More information

Privacy Policy

We use analysis methods (e. g. cookies) to measure how often our site is visited and how it is used. In addition, we use cookies to link your page visits and website use with your customer data stored in our CRM system in order to be able to address you individually, i.e. based on interests and usage. We embed third-party content from other providers (e. g. videos) on our site. We have no influence on further data processing and any tracking by the third party provider. In this context, we also use service providers in third countries outside the EU that may not have an adequate level of data protection, which harbors the following risks: Access by authorities without informing the data subject, no data subject rights, no legal remedies, loss of control.
With your settings you consent to the processes described above. You can revoke your consent with effect for the future. You can find more information in our privacy policy.